products-img

Special Analyzers And Instruments

To protect the world's most vital resources requires instruments, accessories and software that can not only meet but also exceed your productivity, accuracy and regulatory requirements.

Special Analyzers And Instruments

img-prod img-prod img-prod
img-prod img-prod img-prod

TOC/TN pharma

The multi N/C pharma offer tailored solutions for TOC cleaning validation, extractables testing from packaging materials, ultrapure water control, and total protein analysis in aqueous pharmaceutical solutions e.g., vaccines.

Manufacturer: Analytik Jena
Model: multi N/C 3100

Specification:

Models: multi N/C 3100 pharma - TOC/TN analyzer for pharmaceutical applications

Measuring Time: 3 - 5 min

Digestion Mode: High temperature combustion up to 950 oC

Precision: 1 - 2 %

Injection Volume: 100 - 1000 µL

Detector: Focus Radiation NDIR

Sample Supply: Flow injection

img-prod
img-prod

AirComp

Designed for GC and Atomic Absorption, Provides 20 Nl/min of compressed from an 8 bar Oil Free compressor, Possible parallel connection up to 4 stackable tower configuration generators.

Manufacturer: CLAIND

specifications:

FLOW RATE: 20 Nl/min ± 2 Nl/min up to 2000 m from the sea

TECHNOLOGIES EMPLOYED: Oil free piston compressor

OUTPUT PRESSURE: 0÷8 bar ± 3% (± 240 mbar)

AIR OUTPUT CHARACTERISTICS:
. Particulate = 0,01 mm
. Moisture : -15o C dew point ATP

OPERATING TEMPERATURE: 5 oC - 40 oC/40 oF - 100 oF

ENVIRONMENTAL HUMIDITY: max 90% without condensation

img-prod
img-prod

H600, hydrogen Designed for Carrier Gas

Designed for Carrier Gas in GC, GC-MSD and"FAST GC",Safety system with acoustic alarms, screen messages and automatic production stoppage in case of gas leaks, Hydrogen production in proportion to consumption, with minimum gas storage inside the instrument

Manufacturer: CLAIND IT
Model: H600

specifications:

FLOW RATE: 600 Nml/min ± 10% a 2000 m slm

TECHNOLOGIES EMPLOYED: HT PEM electrolytic cell

OUTPUT PRESSURE: 2 - 8 bar ± 3% /30 - 116 psi ± 3%

PURITY: 99,9999 %, Oxygen less than 1 ppm

MOISTURE: Better than -70o C dew-point ATP at 20 oC

REQUIREMENTS OF DEMINERALIZED WATER:
. Required resistivity >1 MΩ cm @25 oC; ASTM Type II
. Advised resistivity >10 MΩ cm @25 oC; ASTM Type I

OPERATING TEMPERATURE: 5 oC - 35 oC / 40 oF - 95 oF

ENVIRONMENTAL HUMIDITY: max 90% without condensation

img-prod
img-prod

Nitrogen, NiGen GC/HC

Designed for GC, Capable of supporting up to 12 make up FID/ NPD, up to 7 ECD or 5 S/S, Ready to be connected to the dedicated AirComp compressor

Manufacturer: CLAIND IT
Model: NiGen GC/HC

specifications:

FLOW RATE: 500 Nml/min ± 10% a 2000 m asl

TECHNOLOGIES EMPLOYED: PSA CLAIND

OUTPUT PRESSURE: 0 - 6 bar/0 - 87 psi ± 3%

PURITY: 99,9995 %

MOISTURE: -50 oC dew-point ATP ± 2 oC ATP

ENVIRONMENTAL HUMIDITY: max 90% without condensation

DIMENSIONS:
. Height: 40 cm + 7 cm/15.8" + 2.8"
. Width: 38 cm/14.9"
. Depth: 53 cm/20.8"